| Throughput | Constant 180T/h (non-ISE) | ||
| Theory | Absorption spectrometry, transmission turbidimetry | ||
| Methodology | End-point Fixed-timeKineticIon selective electrode | ||
| Test according set up, random access | |||
| Reaction plate | Contain 81 reaction cuvettes | ||
| Maximum reaction time | 15 minutes | ||
| Minimum reaction volume | 150 ul | ||
| Material of cuvettes | Special UV-transparent optical glass | ||
| Cuvettes use | Automatically cuvettes washing | ||
| Optical Unit | |||
| Light Source | Halogen-tungsten lamp | ||
| Photometer | Reversed optics | ||
| Wavelength | 340nm-900nm wavelength option | ||
| Absorption | 0-4 Abs | ||
| Reagent Unit | |||
| Reagent plate | 60 reagent positions ,support for single and double reagent, reagent completely open | ||
| Reagent storage temperature | 2~10 degree | ||
| Reagent bottle | 15ml, 20ml, 30ml, 50ml or 70ml ,dead volume is less than 1.0 ml | ||
| Reagent volume | 20-500 ul, step by 1.0 ul | ||
| Reagent/Sample Probe | Automatically checked ;Patent technology liquid level detection;Collision protection probe cleaning:Automatic washing for both inside and outside. | ||
| Sample Unit | |||
| Sample plate | 60 sample positions | ||
| Sample type | Human (animal) serum plasmaurine etc. | ||
| Samples cup | Support one-off micro cups | ||
| Sample volume | 2-100 ul, step by 0.1 ul | ||
| Sample/reagent probe | Automatically checked ;Patent technology liquid level detection;Collision protection probe cleaning:Automatic washing for both inside and outside. | ||
| Temperature control Unit | |||
| Heating method | Direct thermal solid thermostat closed | ||
| Temperature equilibrium time | <3 minutes | ||
| Reaction temperature | 37 ±0.1 degree | ||
| Temperature fluctuation | ±0.1 degree | ||
| Operation Unit | |||
| Software | WindowsXP 2000 | ||
| Database | SQL Server | ||
| Data Storage | Automatic store & copy; No time and volume limit; Lifetime preservation | ||
| Print format | Can be set up, offer several standard format | ||
| Calibration mode | Linear curve type: Single-point linear, two-point linear and multi-point linear none linear; Curve type: Logistic-Log 4P,Logistic-Log5p,Spline,index automatically use best calibration curve and standard | ||
| Quality control methods : | Real time, within day & between day | ||
| Quality control diagram | L-J cumulativetwin plot | ||
| Status display | Real time monitoring of reagent residual volume .Running status of reagent tray, samples tray | ||
| Adjustment | Dynamic and real time display of running status; Intelligent carry over settings to adjust test sequence; original reaction data record; Detail profile of alert messages | ||
| Setting | Operation permission setting, Item setting, cross pollution setting, cuvette blank setting | ||
| Others | |||
| Water consumption : < 6L/h | Power Supply :AC 220V±22V 50Hz±1Hz | ||
To request a quotation for Medical Hospital Laboratory Clinical Chemistry and Biochemistry Analyzer (Product Code HOSPITALLAB70004), share your tender BoQ or requirement list by:
Quotation turnaround is typically 24-48 hours with brochures, datasheets and a priced compliance statement.
As one of India's leading Medical Hospital Laboratory Clinical Chemistry and Biochemistry Analyzer manufacturers, suppliers and exporters, we deliver tender-grade quality with documented compliance to institutions, laboratories and training facilities across 40+ countries.
Every unit is manufactured to recognised international quality standards and tested before dispatch. Full QC documentation and calibration certificates included.
Established export channels across Africa, Asia, the Middle East and beyond. Complete shipping documentation — certificates of origin, packing lists, test certificates, customs paperwork.
Need a specific configuration of Medical Hospital Laboratory Clinical Chemistry and Biochemistry Analyzer? We offer customisation for tender specifications, OEM branding, voltage variants and bulk institutional orders.
Technical consultation before purchase, installation guidance, operator training documentation, and responsive after-sales support — backed by our engineering team.
Common questions from buyers, procurement officers and technical evaluators about ordering, customisation, lead time and documentation.